Features and Benefi

 

High Sensitivity to Helium - Can detect very small leaks

High Sensitivity (2 ppm) to helium, three orders of magnitude better than industry standard,

due to SIPD (proprietary and patented Selective Ion Pump Detection)

Excellent selectivity for helium allows you to read helium leaks and ignore all other gases

Helium sensitivity can be adjusted as required to minimize test costs and helium consumption

Autozero function allows leak detection even in high helium background environment

 

Easy to Use - No training required

State-of-the-art microprocessor control allows great simplicity of operation

Fully automatic start-up

Ready for test in less than 3 minutes

ntuitive touch screen display

Visual and audio indicators (standard headphone connection)

 

No tuning required

 

Truly Portable - Compact and light

The PHD-4 weighs only 2,6 Kg (5.7 lbs) including the battery

Its compact size allows it to be easily carried anywhere* Its ergonomic design allows

comfortable use for extended periods

 

Versatile - Suitable for many different applications

Wide range of uses: replaces or can be used with existing methods such as

 

bubble test or pressure decay

Able to detect both very small and large leaks

Can operate either on battery power or connected to an outlet

Displayed messages can be viewed in several languages (English, French, German, Italian)

Standard Analog and RS232 Serial output.

 

Dependable - Long term operation

Automatic backflow valve prevents helium saturation,

 

ensuring fast response times as well as long life of sensing element.

CE, CSA/US approved for global standardization.

 

The New PHD- 4 Portable Helium Detector

 
 

Technical Specifications

 

 

Lowest Detectable Helium Concentration :

 

2 ppm (parts per million)

 

Lowest Detectable Helium leak :

 

5 x 10-6 mbar l/s

 

 

 

5 x 10-6 atm cc/s

 

 

 

5 x 10-7 Pa m3/s

 

Response Time :

 

< 2 sec

 

Response Time :

 

<10 sec (from 50 ppm to 0 ppm)

 

Start up time, including self check-up :

 

3 min approx.

 

Power Supply :

 

* 12 Vdc, 1.2 A Rechargeable Battery included

 

 

 

* 110-240 V 50-60 Hz Transformer/

     

Battery Charger included

 

Battery operation Time :

 

4 hours

 

Maximum Signal Drift :

 

10 ppm/10 min

 

Operating Conditions Temperature :

 

+5°C to +35°C

 

Humidity :

 

90% maximum relative humidity

 

Storage Conditions Temperature :

 

-20°C to +60°C

 

Weight :

 

2,6 Kg (5.7 lbs)

 

Compliance to Norms :

 

CE approved CSA/US approved

 

Ordering Information

 

 

PHD-4 Complete Package

 

Part Number

 

Travel Case includes

 

9694640

 

* PHD-4 Basic Unit

 

 

 

* Spare Battery

 

 

 

* Transformer/Battery Charger (110-240V)

 

 

 

* Carrying Strap

 

 

 

* Probe Set

 

 

 

* 15-pin I/O connector

 

 

 

* CD Instruction Manual

 

 

 

PHD-4 Replacement Part Kit

 

Part Number

 

Includes

 

9694660

 

* Sampling Pump with Fittings

 

 

 

* Probe with Sampling Line

 

 

 

* Tip Probe Filter

 

 

 

* Internal Filter (Kit of 5 units)l

 

 

 

PHD-4 Basic Package

 

Part Number

 

Includes

 

9694600

 

* PHD-4 Basic Unit

 

 

 

* Transformer/Battery Charger (110-240V)

 

 

 

* Carrying Strap

 

 

 

* 15-pin I/O connector

 

 

 

* CD Instruction Manual

 

 

 

Accessories

 

Part Number

 

* Probe Set

 

9693515

 

* Capillary leak with refillable reservoir and gauge

 

9693540

 

* Probe with 10 meter (30') maximum Sampling Line

 

9693525

 

* Telescoping Extension Probe

 

9693520

 

Individual Replacement Parts

 

Part Number

 

 

 

SR 03.702609

 

* Spare Battery

 

SR 03.702888

 

* Transformer/Battery Charger (110-240V)

 

SR 03.702513

 

* Sampling Pump with Fittings

 

SR 03.702538

 

* Probe with Sampling Line

 

SR 27.590006-01

 

* Tip Probe Filter

 

SR 28.900012-01

 

* Internal Filter (Kit of 5 units)

 

SR 03.702791

 

* Carrying Strap* 15-pin I/O connector

 

SR 03.702894

 

* Travel Case

 

SR 03.702890

 

Applications

 

Large Vessels and Bioreactors

The PHD-4 offers unmatched accuracy and repeatability, presenting a unique solution that it is cost effective and very well suited for the leak range specifications of this application. Biotech and pharmaceutical industries used to rely on pressure decay and bubble test methods for finding leaks in their large bioreactors. The PHD-4 has established a new standard of quality, significantly increasing production yields.

* Fermenters
* Sterilizers
* Freeze Dryers

 

Underground Pipes and Storage Tanks

The portability and light weight of the PHD-4 plays a major role in this application. Underground pipes and storage tanks (UST) are slightly pressurized with helium which, due to its high mobility, can escape through small leaks and migrate to the surface, where it can be easily detected by the PHD-4. The accuracy, portability and light weight of this unit greatly simplifies this process, particularly in difficult construction sites or rough terrain.

* Gas distribution lines
* Under and above ground containers and storage tanks
* Telecommunication and high voltage underground cables

 

Water Heating and Cooling Pipes

The PHD-4 allows leak location without interruption of the normal operation, by mixing helium with the water in the circuit. Until recently, the precise and rapid location of leaks in buried pipes has been very difficult.In the event of a leak, helium desorbs from the fluid and diffuses to the surface, where it is easily detected. Leaks in pipeline systems such as district heating systems, drinking or chilled water systems and steam pipe networks incur high costs due to losses and corrosion damage.

* Heater exchangers and steam condensation lines
* Water pipes
* Radiant heating systems
* Brine pipes

 

Airplane Fuel Tanks and Lines

PHD-4 technology is approved worldwide by airplane manufacturers and operators as the standard for the location of leaks in aircraft fuel tanks and in oxygen distribution lines. For this purpose, helium is injected from the outside into the tank to be checked and the helium probe is used inside the wings to locate any leaks.

* Fuel tanks
* Oxygen distribution lines

 

Other Applications

The PHD-4 is in daily use in many other applications. Virtually any object requiring any level of leak tightness can be simply tested with this instrument. Here is a partial list of other applications:

*Components and systems for the Chemical and Petrochemical Industries
* Components for the Automotive industry
* Process gas delivery lines in Semiconductor fabrication industry